This is actually a feature, not a bug.
Since galvo/galvo scanning is slow, we implemented a feature called 'striping'.
We update a frame stripe by stripe, where one stripe is composed of multiple lines.
This ensures you see the data as soon as it is acquired and you don't have to wait to the end of the frame.
You can configure this feature by manually adding the following lines to the %%LinScan section of the Machine Data File to overwrite the default values:
stripingEnable = true;         % enables/disables striping display
 stripingMaxRate = 10;         % [Hz] determines the maximum display update rate for striping
Usually the image would be chopped up into about 5 stripes, depending upon on your rate (e.g. sample rate is high, you might only get 2 stripes. This is based on stripingMaxRate.
This might look a bit odd on the display.
You can try to disable striping altogether or you can modify the stripingMaxRate to see more/less stripes.
However, don't increase beyond 20Hz; this might result in performance issues.