Scanimage 2017 : ScanImage January 2016 Release Thorlabs BScope 2 Compatibility Information

Affected Users


If you use a Thorlabs BScope2, changes need to be made to your Machine Data File (MDF). This applies to you if you have the following entry in the "ScanImage" section of your MDF:

components = {'ThorBScope2'};

If 'ThorBScope2' is an entry in the components option, follow the instructions below to upgrade your configuration.

Update Instructions


  1. Replace 'ThorBScope2' with 'dabs.thorlabs.BScope2' in the components list.
  2. Update your stage controller settings:
    1. If you have a Thorlabs MCM5000 stage controller:
      1. In the "Motors" section of your MDF, change the motorControllerType from '+ThorBScope2' to 'thorlabs.mcm5000'
      2. Set motorCOMPort to the correct COM port number. This should currently be in the "Thorlabs BScope2" section as "mcm5000ComPort"
    2. If you have a Thorlabs MCM3000 stage controller:
      1. This stage was previously unsupported but now is compatible under this release. In the "Motors" section of your MDF set motorControllerType to 'thorlabs.mcm3000'.
      2. Set motorCOMPort to the correct COM port number.
  3. Find the "Thorlabs BScope2" section of your MDF
    1. Rename "ecu2ComPort" to "ecuComPort"
    2. Remove "mcm5000ComPort" and "hasRotation" fields. They no longer apply
    3. Add the following to the section:

      galvoGalvoMirrorComPort = [];       % Numeric: Serial COM port for controlling Galvo-Galvo mirror. Leave empty is mirrors are connected to stage controller (MCM5000)
      galvoResonantMirrorComPort = [];    % Numeric: Serial COM port for controlling Galvo-Resonant mirror. Leave empty is mirrors are connected to stage controller (MCM5000)
      flipperMirrorComPort = [];          % Numeric: Serial COM port for controlling Flipper mirror. Leave empty is mirrors are connected to stage controller (MCM5000)
      pmtDevAdresses = {[] [] [] []};     % Numeric serial number or string VISA addresses of USB PMT controllers. For each of 4 PMTs leave empty if controlled by ECU or enter addr
      1. USB mirror controllers were previously unsupported. If you have USB mirror controllers, set the COM ports for which ever mirrors apply to your bscope.
      2. For upgrades leave pmtDevAdresses empty. If using any of the new Thorlabs PMT2100 PMT controllers, enter the 5 digit serial number for the controller for each channel.

If you have any questions about these instructions send an email to support@vidriotech.com.