Integration Mode | Photon Counting Mode |
---|---|
![]() | ![]() |
This GUI is opened from the Main Window menu. View > Photon Counting Controls
See Photon Counting for more information.
Photon Counting Controls
Uniform Pixel Bin Size | Disables the geometry correction for the non-uniform resonant scanning speed (see Line Formation) |
Photon Counting Controls - Integration Mode
Differentiate Before Integration | Apply centered differentiation on signal for photo current integration |
Absolute Value Before Integration | Take the absolute value on signal for photo current integration |
Enable Integration Threshold | Ignore any values below threshold for photo current integration |
Integration Threshold | Value of integration threshold (in ADC units) |
Disable Mask Averaging | Disables the division after sample binning (see Line Formation). |
Photon Counting Controls - Photon Counting Mode
Configure Photon Discriminator | Shows the photon discriminator GUI |
Show Photon Histogram | Shows the photon histogram |
Photon Selection Mask | For time correlated photon counting; Binary mask that selects photons after laser pulse used to display in channel |
Disable Mask Averaging | Disables the division after sample binning (see Line Formation). Should be ON for photon counting. |
Photon Discriminator
Grab Data | Grab raw digitizer data. (Sync) |
Save Configuration | Save the current photon configurator settings to disk |
Load Configuration | Load photon configurator settings from disk |
Signal Conditioning | Click into box to open configuration dialog |
Scale by Power of 2 | Scale signal by 2^x |
Offset | Offset signal by x |
Differentiate | Apply centered differentiation filter (0: disabled, 1: enabled) |
FIR Filter Enable | Enable/Disable the FIR filter |
Filter Coefficients | Click in FIR filter plot to open dialog. Maximum of 32 taps, filter values must be integers |
Photon Discriminator | Click into box to open configuration dialog |
Peak Threshold | Threshold for photon detection |
Peak Window | Window size (in samples) for peak detection, must be odd number |
Debounce | Debounce samples |
Plots | Mouse wheel zooms, drag with mouse to pan |
Yellow threshold line | Drag with mouse to set photon peak threshold |
Photon | Click on detected photons, drag window and debounce line to adjust values |