ScanImage acquires voltage signals on each of the enabled input channels. Ideally, the voltage value of 0 would correspond to case of zero signal, i.e. zero detected photons. In reality however, the input digitizer channel and connected detector electronics (e.g. PMT amplifiers) combine to exhibit:
- A non-zero offset value
- A distribution of noise values centered on the non-zero offset value
ScanImage can measure and, optionally, subtract/correct these offset values. Measured offset values are stored in the header (metadata) of saved ScanImage TIF images, within the property 'channelOffsets'. If offset correction is applied by ScanImage (or applied by user in post-processing) a histogram of image data will show a symmetric noise distribution centered on zero (with half negative values) and a positive signal distribution.
ScanImage offset measurements are obtained with:
- Scanner parked in its housing
- Shutter closed (if shutter is configured in Machine Data File)
- Beam modulation fully attenuated (if Beam modulation channel(s) are configured in Machine Data File)
ScanImage offset measurements can be obtained at various times:
Manually |
|
Automatically at start of each acquisition (FOCUS, GRAB, or LOOP) |
|