PMT Controllers Settings Panel
PMT Controller Settings
| PMT Name | Enter a name for the PMT. | 
| Device | Select the NI-DAQ device used to control the associated PMT channel. | 
| Gain AO Channel | Select the Analog Output (AO) channel used to control the gain for each PMT. If select "None", the gain functionality is disabled.  | 
| Power DO Channel | Select the Digital port used to switch each PMT on/off. If select "None", the power functionality is disabled.  | 
| Tripped DI Channel | Select the Digital port used to detect the trip status for each PMT. If select "None", the trip detection functionality is disabled.  | 
| Trip Reset DO Channel | Select the Digital port used to reset the trip status for each PMT. If select "None", the trip reset functionality is disabled.  | 
| Min Control Voltage | Enter the minimum analog output voltage on the NI-DAQ board that controls the PMT gain. An entry is required if an Analog Output (AO) channel is selected for Gain AO Channel (above).  | 
| Max Control Voltage | Enter the maximum analog output voltage on the NI-DAQ board that controls the PMT gain. An entry is required if an Analog Output (AO) channel is selected for Gain AO Channel (above). | 
| Max Gain | Enter the maximum display value for the PMT gain. An entry is required if an Analog Output (AO) channel is selected for Gain AO Channel (above).  | 
| Delete | To delete a row, press the "X" on the row you want deleted. |