This is a normal situation and is expected.
There is a certain amount of noise introduced into the system by the PMT power supply that will always be there.
This is ok if the noise is low (as illustrated in the left picture, between -10 and 10), and your image signal values are well above this noise value.
Try changing the voltage on the PMT power supply and observe what happens.