DAQ resourced used by SI5 Galvo/Galvo, grouped by logical device:

Acq Device

AIs: max 4 channels

  • Analog Input Task: PMT offset (buffered, EveryNCallback)
  • Analog Input Task: measure PMT offsets (unbuffered)

Ctl Device

AOs: 2 channels

  • Analog Output Task: XY galvo control output (buffered, EveryNCallback)
  • Analog Output Task: park/point galvos (unbuffered)

Aux Device

Counters:

  • Counter 0: AI sample clock (continuous pulse train) triggered by start trigger
  • Counter 1: frame clock (finite/continuous pulse train) triggered by start trigger
  • Counter 2: line clock (finite pulse train) triggered by frame clock
  • Counter 3: pixel clock (finite pulse train) triggered by line clock

DIs: 3 channels for Start/Stop/Next Trigger Timestamping

  • Digital Input Task: 3 channels for start/stop/next triggering (continuous buffered, EveryNCallback) triggered by frame clock on first frame of AcqMode, then freerunning