ScanImage 2016 : Configuring LinScan (Galvo Galvo scanning)

Icon

To use the LinScan Imaging System it must be defined as a Scanning System in the General ScanImage Settings section of the ScanImage Machine Configuration Editor.

The ScanImage Machine Configuration Editor can be accessed by selecting the Machine Configuration... option in the File Menu on the Main Controls panel.

 

 

Icon

If you have questions about using existing ScanImage 3.8 hardware with ScanImage 2015, review the article ScanImage 3.x upgrade guide. If you need general support with wiring your system, please write an email to support@vidriotech.com

 

ScanImage 2015/2016 supports galvo/galvo scanning through the module 'LinScan'. LinScan replicates most of the features (triggering, timestamping) of the resonant scanning subsystem (ResScan), but utilizes National Instruments multifunction DAQ boards instead of the FlexRIO platform to acquire data and control the microscope. The recommended hardware for LinScan is a simultaneously sampled NI X-series DAQ board (e.g. NI PXIe/PCIe/USB-63xx). For backward compatibility to ScanImage 3.x, legacy hardware (e.g. NI-6110) is supported as well.

The LinScan module is subdivided into three logical devices: Signal Acquisition DAQ; Auxiliary Digital I/O DAQ; and Galvo Position Control DAQ. Settings for these devices are configured in the Scanner Settings for the associated Linear Scanning system (LinScanner, above), in the ScanImage Machine Configuration Editor. The settings map the logical devices onto the physical DAQ devices available in the computer. Multiple logical devices can be mapped onto the same physical. If the logical devices are mapped to separate physical devices, these devices need to be connected to each other via either a RTSI cable or they all need to be installed in the same PXI chassis.

Icon

Please review the NI-KnowledgeBase What is RTSI and How is it Configured? to learn the correct configuration of a RTSI bus in NI-MAX.

Logical DeviceCommentDAQ Requirements
Acquisition deviceUses analog inputs to acquire image data from the PMTsFor multi-channel input, analog inputs need to be simultaneously sampled. Sample rate ideally >= 1.25MHz
Galvo deviceUses analog outputs to control the positions of the X/Y GalvosRequires at least two analog outputs
Auxiliary deviceUtilizes trigger lines and counters to generate clocks, register triggers and generate hardware timestampsNeeds to be an NI X-series DAQ board (63xx)

Example 1: NI-PXIe\USB 6356

 Dev1 (6356)
Acquisition device: AIs simultaneously sampled, sample rate >= 1.25MHzyes
Galvo device: has two analog outputsyes
Auxiliary device: is NI X-series (63xx)yes

Since the NI-6356 fulfills the requirements for all the virtual devices, no additional hardware is required for LinScan in the Scanner Settings:

 

 

The Galvo device name is selected from the Galvo Position Control DAQ dropdown list.

The acquisition device name is selected from the Signal Acquisition DAQ dropdown list.

The auxiliary device name is selected from the Auxiliary Digital I/O DAQ dropdown list.

Example 1

Wiring diagram for a NI PXIe\USB-6356.

Example 2: NI-6110 in combination with NI-6321

Icon

To exchange timing signals, the NI-6110 and NI-6321 DAQ boards need to be connected with a RTSI bus cable.
Please review the NI-knowledgebase What is RTSI and How is it Configured? to learn the correct configuration of a RTSI bus in NI-MAX.

If you connect PCI-611x board with one or multiple X-series boards (PCIe-63xx), make sure to use the 'Extended Length RTSI Bus Cables' to bridge the gap between the long and short boards (RTSI Bus Cable, Product Number 777562-05)

 Dev1 (6110)Dev2 (6321)
Acquisition device: AIs simultaneously sampled, sample rate >= 1.25MHzyesno
Galvo device: has two analog outputsyesyes
Auxiliary device: is NI X-series (63xx)noyes

 

This allows two possible configurations in the Scanner Settings section of the ScanImage Machine Configuration:

 

                                   Configuration 1                                                                                              Configuration 2

                   

 

The Galvo device name is selected from the Galvo Position Control DAQ dropdown list.

The acquisition device name is selected from the Signal Acquisition DAQ dropdown list.

The auxiliary device name is selected from the Auxiliary Digital I/O DAQ dropdown list.

 

Configuration 1Configuration 2
Signal Acquisition DAQ = 'Dev1'
Galvo Position Control DAQ = 'Dev1'
Auxiliary Digital I/O Daq = 'Dev2'
Signal Acquisition DAQ = 'Dev1'
Galvo Position Control DAQ = 'Dev2'
Auxiliary Digital I/O Daq = 'Dev2'

In Configuration 1, the analog inputs and analog outputs of Dev2 are unused by LinScan, so that Dev2 can also be used as the Beams or FastZ device.

 

Example 2

Wiring diagram for configuration 1. Since the analog inputs and analog ouputs of Dev2 (NI-6321) are unused, they can be configured to control a Pockels cell. See section 'Pockels Control'. 

LinScan Pinout

Acquisition Device

The standard PMT inputs for the Acquisition Device are channel AI 0-3. This configuration can be changed in the Scanner Settings of the ScanImage Machine Configuration.

For instance, to configure the PMT input channels as channels 0,2,4,6 instead of the standard channels 0,1,2,3, select the DAQ Channels as follows:


 

Icon

Please review the article Wiring ScanImage analog input terminals if you are using a NI screw terminal (SCB-68(A) or USB DAQ device) as the breakout box for the Acquisition Device.

Galvo Device

The control output for the X / Y Galvos can be configured in the Scanner Settings of the ScanImage Machine Configuration, by setting the Position Control AO Channel IDs.

 

                      X-Galvo Mirror Channel ID                                                   Y-Galvo Mirror Channel ID

         

 

Icon

A Galvo Position Feedback DAQ can be optionally specified in order to monitor scanning positions.

This can be useful for scanner calibration, arbitrary line scanning and photostimulation feedback.


 

Icon

A Motion Correction Offset DAQ can be optionally specified for motion correctness offset control.

This can be used for motion correction and photostimulation.


 

Auxiliary Device

Outputs

Outputs
Frame ClockPFI10
10MHz Reference ClockPFI14*

The frame clock is output on PFI10.

The 10Mhz Reference Clock is output on PFI14, only if the "Enable export of 10MHz Reference Clock on PFI14" checkbox is checked.


 

Inputs
Start TriggerPFI0 -> port0/line0*
Stop Triggerport0/line1
Next Triggerport0/line2

*PFI0 and port0/line0 need to be bridged to ensure proper time stamping of the start trigger

 

Shutter Control

To configure a shutter for LinScan, review the article Shutter Configuration. If using the Auxiliary DAQ to control the shutter, PFI12 is recommended for the digital output.

Pockels Control

Automatic Trigger Routing

No external trigger wiring between LinScan and the Beams DAQ board is required in the following cases

  1. The LinScan Auxiliary device and the Beams device are configured to operate on the same DAQ
  2. The LinScan Auxiliary DAQ and the Beams DAQ are installed in the same PXI chassis
  3. The LinScan Auxiliary DAQ and the Beams DAQ are connected via a RTSI bus, and the RTSI bus is correctly set up in NI-MAX

For the three cases above, all triggers are routed automatically.

The settings for the automatic routing of triggers are accessed in the Power Modulation (Beams) section of the ScanImage Machine Configuration Editor.

In this case, select "Automatic Routing" for Line Clock, Frame Clock and Reference Clock input terminals:



 

Icon

Please review the NI-knowledgebase What is RTSI and How is it Configured? to learn the correct configuration of a RTSI bus in NI-MAX.

Icon

The Pockels Cell control task is a timed Daqmx Task. Due to Daqmx limitations on timing resources this requires the Pockels Cell to run on a Daq card by itself. Other timed tasks such as Galvo control or FastZ piezo control must run on separate Daq boards.

Automatic Beams Trigger Routing

Trigger Routing via RTSI bus: frame clock and reference clock are automatically routed over RTSI

Icon

Please review the article Wiring ScanImage analog input terminals if you are using a NI screw terminal (SCB-68(A) or USB DAQ device) as the breakout box for the Acquisition Device.

Manual Trigger Wiring

In case the LinScan Auxiliary device and the Beams device are not connected via the PXI/RTSI bus, the LinScan frame clock and reference clock need to be wired to the Beams device.

TriggerSourceDestination
LinScan Frame ClockLinScan Auxiliary board PFI10Beams board any free PFI (e.g. PFI12)
LinScan Reference ClockLinScan Auxiliary board PFI14Beams board any free PFI (e.g. PFI0)

 

The settings for the automatic routing of triggers are accessed in the Power Modulation (Beams) section of the ScanImage Machine Configuration Editor.

In this case configure the settings:

  • Set Line Clock Input Terminal to "Automatic"
  • Set Frame Clock Input Terminal to "PFI12"
  • Set Reference Clock Input Terminal to "PFI0"


 

Manual Beams Trigger Wiring

External Trigger Wiring for Pockels Control

Icon

Please review the article Wiring ScanImage analog input terminals if you are using a NI screw terminal (SCB-68(A) or USB DAQ device) as the breakout box for the Acquisition Device.

FastZ (Piezo) Control

Automatic Trigger Routing

No external trigger wiring between LinScan and the FastZ DAQ board is required in the following cases

  1. The LinScan Auxiliary device and the FastZ device are configured to operate on the same DAQ
  2. The LinScan Auxiliary DAQ and the FastZ DAQ are installed in the same PXI chassis
  3. The LinScan Auxiliary DAQ and the FastZ DAQ are connected via a RTSI bus, and the RTSI bus is correctly set up in NI-MAX

For the three cases above, all triggers are routed automatically so that the FastZ section in the Machine Data File can be configured as follows:

For the three cases above, all triggers are routed automatically.

The settings for the automatic routing of the frame clock trigger for Fast Z is accessed in the FastZ Actuator Setup section of the ScanImage Machine Configuration Editor.

In this case, select "Auto" for the Frame Clock Input Terminal:

 

 

Icon

Please review the NI-knowledgebase What is RTSI and How is it Configured? to learn the correct configuration of a RTSI bus in NI-MAX.

Icon

The FastZ Piezo control task is a timed Daqmx Task. Due to Daqmx limitations on timing resources this requires the FastZ Piezo to run on a Daq card by itself. Other timed tasks such as Galvo control or Pockels Cell control control must run on separate Daq boards.

Automatic FastZ Trigger Routing

Automatic Trigger Routing for Piezo Control

Icon

Please review the article Wiring ScanImage analog input terminals if you are using a NI screw terminal (SCB-68(A) or USB DAQ device) as the breakout box for the Acquisition Device.

Manual Trigger Wiring

In case the LinScan Auxiliary device and the FastZ device are not connected via the PXI/RTSI bus, the LinScan frame clock needs to be wired to the FastZ device:

TriggerSourceDestination
LinScan Frame ClockLinScan Auxiliary board PFI10FastZ board any free PFI (e.g. PFI0)

 

The settings for the automatic routing of the frame clock trigger for Fast Z is accessed in the FastZ Actuator Setup section of the ScanImage Machine Configuration Editor.

In this case, select "PFI0" for the Frame Clock Input Terminal:

 

 

Manual FastZ Trigger Wiring

External Trigger Wiring for Piezo Control

Icon

Please review the article Wiring ScanImage analog input terminals if you are using a NI screw terminal (SCB-68(A) or USB DAQ device) as the breakout box for the Acquisition Device.